== Recommendation ==
Finally, you will probably find it much easier to spend the 7-8 days collecting data at another field strength than to implement the [[#protocol of steps a-i ]] in a relax, Modelfree4, or Dasha script (or via multiple iterations of the GUI programs), as well as study all of the relevant literature to understand all of the types of failures than only occurs with single field strength data. With multiple field strength data you can perform [https://gna.org/users/semor Sebastien Morin's] consistency testing analysis in relax[Morin and Gagné, 2009] (see http://www.nmr-relax.com/manual/Consistency_testing.html). That way you can see if your per-experiment temperature calibration and
per-experiment temperature control techniques have works sufficiently well (http://www.nmr-relax.com/manual/Temperature_control_calibration.html) and if you have used long enough recycle delays. Collecting data at a second field would probably save you significant amounts of time, and has the additional benefit that it would guarantee that the dynamics you see at the end will be real. I cannot emphasize enough how important it is to collect data at multiple fields, most importantly the NOE and R2 data.